Pump Oven

Pump Oven



The pump heads and solvent lines are maintained at a constant temperature within the range of ambient plus 10°C to 50°C, allowing the pumps to deliver exactly the same flow rate under any mobile phase condition. Most HPLC pumps do not compensate for volume changes as a function of temperature. The EcoSEC GPC system has a dedicated oven just for the pumps, thus ensuring the same volume of solvent delivered under any condition.

The first Figure below illustrates the superior baseline stability for the EcoSEC GPC system compared to a conventional GPC system when the temperature in the lab was changed by 3°C every few hours.

The last two Figures demonstrate very low RSD values for retention time and calculated molecular mass when repeated injections were made of narrow molecular weight polystyrene standards and a broad molecular weight vinyl chloride/vinyl acetate/vinyl alcohol terpolymer.


Pump Oven
Temperature control method: Hot-air agitation under PID control
Temperature control range: Room temperature +10 to 50ºC (1ºC steps)
Temperature setting accuracy: ±0.5ºC
Temperature control precision: ±0.1ºC
Heater: 24 VDC, 60 W
Safety mechanism: Temperature sensor for monitoring the in-tank temperature and stopping temperature control upon detection of an abnormality
Gas sensor sensitivity for stopping temperature control
Thermal fuse (70ºC) for shutting down the temperature control circuit
Door open/close monitor for stopping temperature control when the door is opened
Liquid drain in case of liquid leakage
 
Room temperature effect on retention time

Room temperature effect on retention time

Reproducibility of retention time

Reproducibility of retention time

Reproducibility of molecular weight

Room temperature effect


 

 


 